Department: Quality Improvement Document Identifier: GL/QID/1001/01
This guideline specifies the process to be observed within the associated guidelines regarding the management of medical equipment in a clinical environment. In addition, all relevant rules, codes, regulations instructions or other legislation, adopted by the guideline are identified.
This guideline has been developed to ensure that all Healthcare Facilities are adequately and consistently evaluated for compliance with the standards provided within this guideline.
| 1. PURPOSE | |
|---|---|
| 1.1 | To govern the management of medical equipment in healthcare facilities licensed by DHCA. |
| 1.2 | To comply with the statutory authorities and associated guidelines described herein, all healthcare facilities should have at least reliability and redundancy in the provision of essential medical equipment. Without such provisions, proprietors risk their duty of care to patient safety associated with operating health facilities. |
| 2. SCOPE OF APPLICATION | |
|---|---|
| 2.1 | This guideline applies to all healthcare facilities licensed by DHCA. |
| 3. GUIDELINE | |
|---|---|
| 3.1 | Healthcare facilities should have a plan for maintaining and managing medical equipment |
| 3.2 | This plan should include the management of the following: 3.2.1 Medical Equipment Management 3.2.2 Equipment Acquisition 3.2.3 Equipment Tagging 3.2.4 Breakdown Maintenance 3.2.5 Preventative Maintenance 3.2.6 Medical Device Recall 3.2.7 Safety Testing 3.2.8 Management of Compressed Gas |
| occurs, maintenance responsibilities, and the reporting procedures for equipment problems, failures and user errors. | |
| 3.5 | Medical Equipment Plan |
| 3.5.5 | Hazards Notices & Recalls: 3.5.5.1 All electrical patient care equipment should be evaluated based on function and physical risks associated with patient use. 3.5.5.2 All product safety alerts, hazard notices, and/ or recalls will be directed to the facility management. 3.5.5.3 In the event equipment must be removed from service, the equipment should be replaced with a safe and effective substitute. 3.5.5.4 The facility should tag and remove any defective equipment from use due to any recall notices until it can be rendered safe. |
| 3.5.6 | Performance Evaluation: Healthcare facilities should keep records of equipment performance and identify quality indicators that cove preventive & breakdown maintenance, incidents etc. |
| 3.6.2.3 | A copy of the manufacturer's operating manuals and instructions must be made accessible to staff to review at the location of the equipment. |
| 3.7 | Medical Equipment Tagging system |
| 3.10.3.3 Equipment is to be removed from the use and informed to the supplier/manufacturer. | |
| 3.10.3.4 Reports are to be documented and kept in the equipment file. |
or being knocked over. Suitable racks, straps, chains or stands or other devices should be required to support cylinders.
3.12.2.7 Cylinders should not be exposed to temperature extremes nor stored in the vicinity of combustible substances. 3.12.2.8 MRI compatible cylinders should be placed in the MRI area and no other cylinder should be around the area. 3.12.2.9 Never use a cylinder without a regulator. Always use the correct pressure regulator. After attaching the regulator and before the cylinder is opened, check the adjusting screw of the regulator to see that it is released. Never permit the gas to enter the regulator suddenly. 3.12.2.10 Never strike an electric arc on a cylinder. Never use a damaged cylinder. Never force a cap or regulator. 3.12.2.11 Avoid dropping and striking cylinders together. Avoid dragging, sliding or rolling cylinders and always use trolleys for transportation. A sign should be displayed on doors stating "No Smoking". Storage rooms should be dry, cool, and well ventilated. 3.12.2.12 Limited cylinders should be stored in patient care areas. When small-sized cylinders are in use; they should be attached to cylinder stand. An individual cylinder placed in patient room for immediate use by a patient should not be required to be stored in an enclosure but in cylinder trolley with lock. Cylinders should not be chained to portable or movable apparatus. Cylinders should not be stored near elevators, or in corridors. Cylinders should be protected from tampering by unauthorized persons and will be kept capped at all times. Free standing cylinders should be properly chained or supported in a proper cylinder stand or cart. 3.12.2.13 In the event of the leak or suspected leak of a toxic or flammable gas, evacuate from the building or the area. Activate the fire alarm by pulling the nearest break glass unit (B.G.U) and immediately call health & safety department to initiate the proper code. The event is to be reported to DHCR HSE Department in addition to Security for oversight and investigation (where appropriate). 3.12.2.14 "No Smoking" signs should be placed near flammable gas cylinders. Fire-suppression equipment using carbon dioxide or dry chemicals should be available. 3.12.2.15 "No Smoking" signs should be placed near flammable gas cylinders. Fire-suppression equipment using carbon dioxide or dry chemicals should be available.
| 3.1 | References to Licensed Healthcare Professionals means those Healthcare Professionals licensed or certified under the Healthcare Professionals Regulations. |
| 3.2 | “Accreditation body” or “body” means an entity that has been approved by FDA to accredit mammography facilities. |
| 3.15 | “Manager” means the person who performs those functions with respect to a Healthcare Operator that are described for a Manager in the DHCC Healthcare Operator Regulations. |
| 3.16 | “Medical Equipment” means machinery designed to aid in the diagnosis and treatments of medical problems with rigorous safety standards. |
| 4.1 | Appendix 1- SOP for Medical Gas Outlets |
| 5.1 | Health Technical Memoranda (Institute of Electrical and Electronics Engineers (IEEE) |
| 5.2 | Medical Equipment Management Manual by (AAMI) |
This is inclusive of Medical Compressed Air, Nitrogen, Nitrous oxide, Oxygen, Carbon Dioxide and other gases if piped.
| Medical Gas | Pressure in psig | Flow l/min |
|---|---|---|
| Oxygen | 50-55 | > or = 100 |
| Nitrous Oxide | 50-55 | > or = 100 |
Open the flow control valve until a flow of 100 l/min is seen. Pressure at the outlet should not dip below 50 psig for all gases except Nitrogen, which should not drop below 160 psig at a flow of at least 145 l/min. However, if the pressure drops to below 80% of the listed values in the older systems or below the required values in the new systems, corrective action is required, Unacceptable pressure or flow may indicate a blockage in either the distribution piping or the outlet check valves.
If the above issues are not resolved, the outlet may have to be replaced or depending on the severity of the restriction, a portion of the system may have to be modified. Please also consider whether simultaneous use of multiple outlets will further degrade performance.
In some hospitals, hoses extend from ceiling connectors to outlets that are suspended at a lower more accessible height. Although the ceiling connector and suspended outlets may have proper labels and unique fittings for each gas to prevent incorrect connections, the end connections of the hoses and the pipelines to the gas fittings may be identical. Thus it may be possible to attach an outlet or connector to the wring hose. If you have such an installation, please make sure that the ceiling connector and outlet linked by a given hose have the same gas label. The problem may arise if your use extension hoses to connect equipment to a wall or ceiling outlet.
Gas hoses should have appropriate connectors for attachment to equipment. Avoid using special adaptors for connecting hoses to minimize problems such as gas leaks at the connectors. Color coded hoses are recommended for this application to reduce risk or misconnection.